IntroductionParameter


| Sensor type | 4 wires / 2 wires PT-100 |
| - flexible | 4 wires: 480 mm (19”) / 2 wires: 150 mm (6") or custom length |
| sensor size | 4 wires: 1.7 x 10 mm (0.07 x 0.39”) |
| cross section |
2 wires: 1.2 x 2.8 mm (0.05 x 0.11”) 4wires: 2.27 mm2 (0.0035 inch2) 2wires: 1.13 mm2 (0.00175 inch2) |
| Operating Range | |
|
- body temp. - 4 wires temperature. - 2 wires temperature. |
-90°C至+140°C -200°C 至 +200°C -90°C 至 +140°C |
| Accuracy | |
|
- 4 wires - 2 wires |
±0.1°C from -90 to +140°C ±0.2°C from -90 to+140°C |
| Resolution | 0.01°C |
| Memory capacity | 32000 data points |
| Sampling rate | 1 sec to24 hours |
| Transmission | Simultaneous: maximum 20 loggers Sequential: unlimited |
| Internal clock Drift | 4 sec / 24 h @ + 23 °C |
|
Mechanical - material - enclosure - weight - diameter - height |
Peek 1000 - FDA compliant IP-68 50 g (1.76 oz) 37 mm (1.46”) 62.81 mm (2.47”) |
| Power |
2 x Lithium batteries, 0.35 g/each Lithium content classified UN3091 under IATA DGR section 3.9.2.6 compliant with ICAO/IATA packing instruction 970 user replaceable |
| Battery lifetime | Up to 6 months, based on usage |
| Antenna | Internal |
| Wireless range | 121 m (400’) |
| Radio frequency | 2.4 GHz ISM band |
| Access point | USB to computer |
| Calibrations |
Factory:Traceable NIST / COFRAC - ITS-90 coefficients reside in the internal memory User:Closed loop user calibration, using XpertLog® Software |
| Certification |
EMC:EN 60601-1-2 / 2007 RF:EN 300440-2 V1.3.1(2009—03) Health & Safety: EN 60950-1/2006 |